International standard IEC61000-4-11(2004) Correspondence voltage descent examination device
Enable to dip, short time black out, voltage change examination.
Correspondence to single phase(100V) - 3Phase(440V) examination.
The correspondence manipulate is possible from touch panel, external PC and RS232C.
Award to the Saitama Industry personal club (Nishikaiki commemoration prize)!
| Input voltage (A transformer require 3phase 4wire type if an input voltage is 3phase 3wire type.) |
3φ4W,50/60Hz | Between L to N:100V - 254V (-10%,+10%) |
|---|---|---|
| Between L to L:173V - 440V (-10%,+10%) |
||
| Out put voltage | Main powers 100 - 5% | A slide transformer can manipulate serial variable, 0% is short reason to electromagnetic contactor. |
| Output current | Max 30A (serial rating) | An examination characteristic of IEC is regulation to 16A. |
| time rated value | 80%output:30A Concatenation→20A 5sec in IEC standard | |
| 70%output:30A Concatenation→20A 5sec in IEC standard | ||
| 40%output:40A 3min | ||
| Load change | 100%output:0 - 16A less than 5% | |
| 80%output:0 - 20A less than 5% | ||
| 70%output:0 - 23A less than 5% | ||
| 40%output:0 - 40A less than 5% | ||
| Peak current provide ability | 100V - 120V:250Amax | |
| 200V - 240V:500Amax | ||
| 250V - 415V:1000Amax | ||
| Over shout・under shout | Constant voltage less than 5% | |
| Voltage change time (moment stop examination item) |
1 - 5μS(100ΩResistance Load time) | |
| Voltage switch starting phase (moment stop examination item) |
0 - 360°(1°step) Phase measurement error: Less than ±10° | |
| Voltage change mode (moment stop・voltage change common examination item) |
Independence the each single phase: Correspondence neuter point: phase space | |
| all phase together : Correspondence neuter point | ||
| Under voltage establishment item (Moment stop examination・Voltage change common examination item) |
100 - 0%(1%step) | |
| Voltage drop time (moment stop examination item) |
0.1 - 250/300cycle(0.1cycle)However less than examination cycle. | |
| Voltage increase time (moment stop examination item) |
moment change and 100/120 - 999cycle(1cycle step) | |
| Descent voltage maintenance time (Voltage fluctuation examination item) |
1 - 9999cycle(1cycle step) | |
| Voltage increase time (Voltage fluctuation examination item) |
25/30 - 9999cycle(1cycle step)But, minimum of configuration time has setting value depends on decrease voltage configuration value.(70%:25/30cycle,100%:80/100cycle) | |
| Examination cycle (moment stop examination item) |
moment examination :250/300cycle - 9999cycle | |
| voltage fluctuation examination:250/300cycle - 9999cycle | ||
| The number of exanimation time (moment stop/voltage fluctuation common exanimation item) |
1 - 99time and concatenation | |
| Save configuration data | within memory | |
| External memory: PC display by RS232C connection | ||
| operation method | panel side: Touch panel | |
| Outside operation:PC display by RS232C connection | ||
| Power control | 1φ2W,50/60Hz,100/110V(-10%,+10%) | |